A7:TG-DTA-MSNanoGREEN Bldg. E-501

|
Maker |
Rigaku |
|---|---|
|
Model |
TG-DTA8122HMSy |
|
Use for |
Simultaneous measurement of the mass change of the material due to thermal load and the mass spectrum of the emitted gas. |
|
Outline |
An apparatus that connects a thermal analysis device capable of measuring mass changes and endothermic/exothermic reactions with a quadrupole mass spectrometer via a capillary tube. |
Specification
|
Measurement method |
Horizontal differential triple coil method |
|---|---|
|
Temperature Range |
Room temperature to 1500°C (normal use 1350°C) |
|
Maximum heating rate |
100°C/min |
|
Maximum sample amount for measurement |
1g |
|
Scale |
TG ±250mg, DTA ±1000μV |
|
Measurement atmosphere |
Helium |
|
Measurable mass range |
m/z=1 to 100 |
|
esolution |
M/⊿M=2M" |
Noticen
- This equipment is installed in the dry room.
- In principle, samples that generate toxic or corrosive gases cannot be measured.