A7:TG-DTA-MSNanoGREEN Bldg. E-501

A7:TG-DTA-MS

Maker

Rigaku

Model

TG-DTA8122HMSy

Use for

Simultaneous measurement of the mass change of the material due to thermal load and the mass spectrum of the emitted gas.

Outline

An apparatus that connects a thermal analysis device capable of measuring mass changes and endothermic/exothermic reactions with a quadrupole mass spectrometer via a capillary tube.

Specification

Measurement method

Horizontal differential triple coil method

Temperature Range

Room temperature to 1500°C (normal use 1350°C)

Maximum heating rate

100°C/min

Maximum sample amount for measurement

1g

Scale

TG ±250mg, DTA ±1000μV

Measurement atmosphere

Helium

Measurable mass range

m/z=1 to 100

esolution

M/⊿M=2M"

Noticen

  • This equipment is installed in the dry room.
  • In principle, samples that generate toxic or corrosive gases cannot be measured.